A new Cr(VI)-selective electrode of PVC membrane based on triheptyl dodecyl ammonium iodide is described. The response of the electrode is Nernstian for Cr(VI) concentration down to 2X10-6M in 0.03-0.13N HF solution. Limit of detection is 5X10-7M. The plot of logarithms of selectivity coefficients vs. z/γ of various anions is essentially linear with a correlation coefficient of -- 0.88. Applications of the electrode in determination of total Cr in steels and copper alloys, and Cr(VI) in waste-water from chromium electroplating are reported.
Key words: ION SELECTIVE ELECTRODE, POLYVINYL CHLORIDE, MEMBRANE ELECTRODES, CHROMIUM ION
CLC Number:
O651
FENG DAMING;CHEN CHENGZHI. A study on ion-selective electrodes based on ion associates I: A new Cr(VI)-selective electrode of PVC matrix membrane[J]. Acta Chimica Sinica, 1983, 41(4): 371-374.