Acta Chimica Sinica ›› 2004, Vol. 62 ›› Issue (16): 1551-1556. Previous Articles     Next Articles

MALDI/FTMS质量校准新方法

苏越1, 陈国强2, 郭寅龙2, 相秉仁1, 安登魁1   

  1. 1. 中国药科大学分析测试中心, 南京, 210009;
    2. 中国科学院上海有机化学研究所上海质谱中心, 上海, 200032
  • 投稿日期:2004-02-01 修回日期:2004-04-20 发布日期:2014-02-17
  • 通讯作者: ylguo@mail,sioc.ac.cn
  • 作者简介:郭寅龙:E-mail:ylguo@mail,sioc.ac.cn
  • 基金资助:
    国家自然科学基金(No.20175034)资助项目.

Novel Approach of Mass Calibration in MALDI/FTMS

SU Yue1, CHEN Guo-Qiang2, GUO Yin-Long2, XIANG Bing-Ren1, AN Deng-Kui1   

  1. 1. Analysis and Testing Center, China Pharmaceutical University, Nanjing 210009;
    2. Shanghai Mass Spectrometry Center, Shanghai Institute of Organic Chemistry, Chinese Academy of Sciences, Shanghai 200032
  • Received:2004-02-01 Revised:2004-04-20 Published:2014-02-17

Because all ions in the same experimental conditions undergo the same shift induced by the electric field associated with the charge of the ions in the analyzer cell in MALDI/FFMS, the constants A and B in calibration equation(m/z=A/f+B/f2)are amended by the shift of matrix peak observed frequency.The mass obtained by the amended calibration equation is therefore immune to space-charge effects on mass accuracy.Comparison of prediction performance of the calibration models constructed by multivariate regression and simplex method was studied.Simplex method, in which maximum relative error acts as objective function, is superior on solidity of ion mass prediction.And the relative errors of ion mass measured all are less than 5x10-6.The IRMPD fragment ions of clarithomycin were determined by this method, and the fragmentation patterns were discussed.

Key words: MALDI, FYMS, mass calibration