化学学报 ›› 1958, Vol. 24 ›› Issue (2): 194-196. 上一篇    下一篇

论文

络合滴定钙镁合量的新指示剂

郭小伟   

  1. 冶金工业部地质局鞍山分局检验室
  • 投稿日期:1957-07-13 发布日期:2013-06-03

BERYLLON Ⅱ AS AN INDICATOR FOR COMPLEXOMETRIC TITRATION

KWAUK SHEAU-WEI   

  1. Laboratory of Anshan Branch, Bureau of Geology, Ministry of Metallurgical Industry
  • Received:1957-07-13 Published:2013-06-03

用络合滴定法测定钙、镁含量时,程氏[1]建议用Calcon的锌盐作指示剂。Стюн-кель[2]用酸性深蓝及酸性蓝,但均无显著优点。本文报告改用Beryllon Ⅱ(变色酸与H-酸偶联而成的偶氮染料)的结果和此指示剂的一些特点。(一)探用新指示剂测定钙镁含量:Beryllon Ⅱ在微酸性中呈红色,而在pH=9.5—10.6时变为紫色。

Beryllon Ⅱ [8-oxynaphthalene-3,6-disulfonic acid-l-azo-2'(-1',8'-dioxynaph-thalene-3',6'-disulfonic acid)tetra sodium salt] is recommended as a substitute for Eriochrome black T in the complexometric determination of the sum of calcium and magnesium,such as in the case of limestone,dolomite etc.At pH=10.0,the end point is easily detected as the color changes from blue-violet to violet-red.Distinct from Eriochrome black T,titanium,iron and aluminum do not form any colored complex at the recommended pH.Therefore,calcium and magnesium can be titrated directly with EDTA in presence of these metals by using potassium-sodium tartrate as masking agent.