Acta Chimica Sinica ›› 2006, Vol. 64 ›› Issue (7): 705-708. Previous Articles    

Reports

微流控芯片电泳快速分离脂蛋白

王惠民*,1,丛辉1,孙承龙1,金庆辉2,贾春平2,宋宏伟1   

  1. (1南通大学附属医院 南通 226001)
    (2中国科学院上海微系统与信息技术研究所 上海 200050)
  • 投稿日期:2005-06-30 修回日期:2005-12-02 发布日期:2006-04-15
  • 通讯作者: 王惠民

Rapid Separation of Lipoproteins Based on Microfluidic Chip Electrophoresis

WANG Hui-Min*,1, CONG Hui1, SUN Cheng-Long1, JIN Qing-Hui2, JIA Chun-Ping2, SONG Hong-Wei1   

  1. WANG Hui-Min*,1, CONG Hui1, SUN Cheng-Long1, JIN Qing-Hui2, JIA Chun-Ping2, SONG Hong-Wei1
  • Received:2005-06-30 Revised:2005-12-02 Published:2006-04-15
  • Contact: WANG Hui-Min

A rapid assay of separating lipoproteins with microfluidic chip is described. The pre-stained lipoproteins with nitrobenzoxadiazole-6-ceramide were separated and analysed in quartz chip with laser induced fluorescence (LIF) detector. The electrophoresis buffer was accomplished with 40 mmol/L tricine buffer by addition of 40 mmol/L methylglucamine. After 40 s sample injection by applying 500 V voltage, the 2000 V separation voltage was switched on, and the peaks of low-density lipoprotein (LDL) and high-density lipoprotein (HDL) were detected within 2 min. After calculation coefficient of variation (CV) of sample migrating time was not more than 2.6% through 5 times detection. The method was sensitive, rapid, simple and reproducible for diagnosis of hyperlipoidemia.

Key words: microfluidic chip electrophoresis, lipoprotein, assay