A STUDY OF ANODIC STRIPPING VOLTAMMETRY OF TRACE GERMANIUM ON GOLD FILM ELECTRODE

DENG JIA-QI1, ZHANG QING-YUAN2, WANG ZHAI-ZHONG3   

  1. 1. Chemistry Department, Fudan University, Shanghai;
    2. Chemistry Department, Anhui University, Hefei;
    3. Chemistry Department, Yunnan University, Kunming
  • Received:1980-10-20 Published:2013-06-03

The anodic stripping voltammetry of trace amounts of germanium on the gold film deposited on a glassy graphite electrode was reported.In boric acid buffer solution, at pH 10, two peaks of stripping current of germanium appear at -0.90 V and -0.17V(vs.1N Ag-AgCl)respeotively, the first peak is used for quantitative analysis.A linear relationship exists between the peak height and the concentration of germanium in the range of 1.0×10-8~1.0×10-6M.The detection limit is about 5.0×10-9M (or 0.36 ppb).We have further studied the electrode process by means of triangle cyclic voltammetry and proved that the electrode reaction is irreversible and show that the first peak is due to the oxidation of metallic germanium to divalent germanium ion and the second peak due to the oxidation of divalent ions to tetravalent ions.The doposition of germanium on the gold film electrode is possibly due to the formation of an alloy of germanium and gold.