化学学报 ›› 1987, Vol. 45 ›› Issue (6): 546-550. 上一篇    下一篇

研究论文

间断多次扫描电位溶出分析方法的研究

方禹之;何品刚;金利通   

  1. 华东师范大学化学系
  • 发布日期:1987-06-15

Study on interrupted multiple-scan potentiometric stripping analysis

FANG YUZHI;HE PINGANG;JIN LITONG   

  • Published:1987-06-15

关键词: 扫描, 镉, 铅, 尿, 电位溶出分析法, 间断, 图盘电极

Interrupted multiple-scan potentiometric stripping anal. is a new technique of potentiometric stripping anal., which is developed from multiple-scan potentiometric stripping anal. and interrupted voltammetry. The technique not only maintains the high sensitivity of multiple-scan potentiometric stripping anal., but also can eliminate the interferences from elements commonly encountered in this method. Experimental conditions are investigated for determination of Cd2+ and Pb2+, which were determine simultaneously by the new technique with a detection limit of 0.005 ng Cd2+/mL and 0.01 ng Pb/mL after 10 min electrolytic deposition. The relative standard deviations are 2.9% and 3.2%, resp., for detns. of 4 ng/mL of Cd2+ and Pb2+. Interrupted multiple-scan potentiometric stripping anal. can compete with the pulse anodic stripping voltammetry in microanal. field.

Key words: SCANNING, CADMIUM, LEAD, URINE, POTENTIOMETRIC STRIPPING ANALYSIS, DISCONTNUITY, DISK ELECTRODES

中图分类号: